DSpace logo

Please use this identifier to cite or link to this item: http://142.54.178.187:9060/xmlui/handle/123456789/19697
Title: Development and Validation of Reversed-Phase High Performance Liquid Chromatographic Method for Analysis of Cephradine in Human Plasma Samples
Authors: MAHMOOD AHMAD
MUHAMMAD USMAN
ASADULLAH MADNI
NAVEED AKHTAR
NAYAB KHALID
WAHEED ASGHAR
SATTAR BAKHSH
Issue Date: 20-Feb-2010
Publisher: HEJ Research Institute of Chemistry, University of Karachi, Karachi.
Citation: Ahmad, M., Usman, M., Madni, A., Akhtar, N., Khalid, N., & Asghar, W. (2010). Development and validation of reversed-phase high performance liquid chromatographic method for analysis of cephradine in human plasma samples. Journal of the Chemical Society of Pakistan, 32(1), 58-63.
Abstract: An HPLC method with high precision, accuracy and selectivity was developed and validated for the assessment of cephradine in human plasma samples. The extraction procedure was simple and accurate with single step followed by direct injection of sample into HPLC system. The extracted cephradine in spiked human plasma was separated and quantitated using reversed phase C18 column and UV detection wavelength of 254 nm. The optimized mobile phase of new composition of 0.05 M potassium dihydrogen phosphate (pH 3.4)-acetonitrile (88:12) was pumped at an optimum flow rate of 1 mL.min -1 . The method resulted linearity in the concentration range 0.1520 µg.mL-1The limit of detection (LOD) and limit of quantification (LOQ) were 0.05 and 0.150 µg.mL-1 , respectively. The accuracy of method was 98.68 %. This method can be applied for bioequivalence studies and therapeutic drug monitoring as well as for the routine analysis of cephradine.
URI: http://142.54.178.187:9060/xmlui/handle/123456789/19697
ISSN: 0253-5106
Appears in Collections:Issue 01

Files in This Item:
File Description SizeFormat 
ViewByVolume.aspx?v=26&i=VOLUME%2032,%20NO1,%20FEB%202010.htm148 BHTMLView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.